ENGLISH


The Klein Tools MM6000 is a True RMS, auto-ranging multimeter.
It measures AC/DC voltage, low impedance voltage, AC/DC current,
resistance, capacitance, frequency, duty cycle, and temperature. It can also
test diodes and continuity.
• 6,562 ft. / 2,000 m
• 75%
• 32°F to 113°F (O°C to 45°C)
• -4°F to 140°F (-20°C to 60°C) <80% R.H.
• 18°C to 28°C (64°F to 82°F)
•  0.1*(specified accuracy)/ °C
• 3 samples per second
•  7” x 3.5” x 1.875” (178 x 89 x 48 mm)
• 14 oz. (397 g)
• 160 Hours
•  3 m
•  20
• Accurate for one year
• ± (% of reading + # of least significant digits)
• CAT III 1000V, CAT IV 600V
• Conforms to UL Stds 61010-1 & 61010-2-030.
Certified to CAN/CSA C22.2 No.61010-1, 61010-2-030,
IEC 61010-2-030:2010 MOD, ANSI/ISA-61010-2-030 (82.02.03).
• 2
• USA (of US and imported parts)

To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring a known
voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep fingers away from the
metal probe contacts when making measurements.
• Do not open the meter to replace batteries or fuses while the probes
are connected.
• Use caution when working with voltages above 60V DC or 25V AC
RMS. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock, replace
batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lock out power
before measuring resistance or capacitance.
• Always adhere to local and national safety codes. Use individual
protective equipment to prevent shock and arc blast injury where
hazardous live conductors are exposed.

~
 
 
 
 
C
E
A
G
F
H
J
I
L
K
N
B
5
8
3
2
 
Features:
RANGEREL


• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• Rotate selector to "AC" Voltage position.
• Attach RED lead to “V” input, BLACK lead to COM.
 
Features:
RANGEREL
 
Features:

• Attach RED lead to “μA” input, BLACK lead to COM.
• Break circuit and connect meter in series (AS ABOVE).
• Select AC or DC current source.
 
Continuity Features:

• Display shows resistance.
• Buzzer sounds if less than 40Ω.
Diode Features:

Display shows:
• Forward voltage drop if forward biased.
• “O.L” if reverse biased.


 RANGEREL
Resistance Features:
•
 measure resistance on a live
circuit.
 Ω = < 50MΩ.

REL
RANGE 

• Start with this setting if current level is unknown.
• Break circuit to be measured.
• Attach RED lead to "A" input, BLACK lead to COM.
• Connect RED lead to one side of break and BLACK lead to the
other (meter in series).
• Meter will automatically select "A" when RED lead is connected to
"A" input. “” will show if connected but fuse is blown.
 
Features:
RANGEREL


• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• Rotate selector to "DC" Voltage position.
• Attach RED lead to “V” input, BLACK lead to COM.

 






 
• Device will power off after 30 minutes non-use.
• Press the select button to wake.
• Disabled during Max/Min function.
• Holding Select button while turning on disables Auto Power-Off.
 
• Switch between AC and DC.
• Switch between “Ω”, “
”, “ ”, and “ ”.
• Switch between Hz and %.
• Switch between °F and °C.
• Wake meter during Auto Power Off.
 
• Press to store current value.
• Display shows the difference between the stored and live readings.
• Press again to return to live reading.
 
• Press repeatedly to cycle through manual ranges.
• Press for 2 seconds to return to auto ranging mode.
• AT is displayed on LCD only during auto ranging mode.
 
• Press to enter Max/Min mode; the largest and smallest values will be
saved while in this mode.
• Press repeatedly to alternate between the maximum and minimum
readings.
• Press for 2 seconds to return to live reading and clear the stored
maximum and minimum values.
 
• Hold captures the current displayed value.
• Auto Hold captures the first stable displayed value, until a new stable
value is measured. The meter will then capture the new value and emit
a beep (V, Ohm, Temperature and Amp functions).
• Press to hold current input on the display.
• Press again to enter Auto-Hold to capture stable values.
• Press again to return to normal operation.
• Press and hold to activate display backlight.
• Using lights drains the battery significantly.
 
• When indicator is displayed on the LCD, batteries must be
replaced.
• Remove rubber boot, back screw and replace 2 x AAA batteries.
• This meter uses 440mA/1000V (69192) and 11A/1000V (69191)
fast blow fuses.
 (optional, sold separately)
• Slide magnetic adapter into protective rubber boot.
• Attach instruments to metal for hands-free use.
 
• The bar graph shows an approximate analog representation
of a measurement.
• The bar graph responds much faster than the digital display.
• The scale of the bar graph is zero to the maximum reading of
the selected range.
